We have many years of production and sales experience, we have accumulated rich strength in the Light Fuel Oil Purifier, kinematic viscometer, Dry Air Generator field. We wholeheartedly provide our customers with corresponding products and support, and look forward to forming a stable relationship with friends from all walks of life and working hand in hand to create a better future! We recruit a variety of professional and skilled personnel, and conduct comprehensive training and examination of employees. With the production and service goal of 'good quality stability, strong price competitiveness and high customer satisfaction', our company is constantly committed to innovating products and looking forward to establishing long-term and mutually beneficial cooperative relations with customers at home and abroad.
1. Introduction & Application
FT-YJS-058A Auto Cleaning Type Insulating Oil Dielectric loss and resistivity tester, is designed and manufactured based on GB/T5654-2007 of (Liquid insulating material of relative permittivity, dielectric loss factor and DC resistivity measurements). It is used for measuring dielectric dissipation factor and DC resistivity of insulating oil and other insulating liquids.
This Oil Tan Delta Test kit is characterized by integrated structure, and integrates test cell, temperature controller, temperature sensor, dielectric loss test bridge, AC test power, standard capacitor, high resistance meter, DC high voltage source and other major components. The transformer oil tan delta test set adopts full digital technology and all of the intelligent automatic measurement, is equipped with large color screen LCD display, whose each step has prompted the English, the test results can be automatically stored and printed output, so that the operator without professional training will be able to use it proficiently.
2. Why Test Oil Dielectric Loss at High Temperature?
High temperature Tan Delta tests are more sensitive to small changes that occur in oil characteristics. Some ionic contaminants may not be detectable at 25°C but are revealed in the 100°C test. Measuring Tan Delta or Resistivity at two temperatures, e.g., ambient and 90 °C (IEC 60247), as opposed to just one temperature, may provide additional, useful information. For example, obtaining a high Tan Delta test result at ambient and an acceptable Tan Delta test result at a high temperature typically indicates the presence of moisture, since the moisture will evaporate at higher temperatures. Conversely, a high Tan Delta test result at both temperatures (or at higher temperatures only) is often attributed to the presence of contaminants.
3. Technical Parameters
Voltage supply |
AC 220V±10% |
Power frequency |
50Hz/60Hz ±1% |
Measuring range |
|
Capacitance |
5pF~200pF |
Relative Permittivity |
1.000~30.000 |
Dielectric loss factor |
0.00001~100 |
DC Resistivity |
2.5MΩm~20TΩm |
Measuring accuracy |
|
Capacitance |
± (1% of measured value+0.5pF) |
Relative Permittivity |
±1% of measured value |
Dielectric dissipation factor |
± (1% of measured value +0.0001) |
DC Resistivity |
±10% of measured value |
Resolution |
|
Capacitance |
0.01pF |
Relative permittivity |
0.001 |
Dielectric dissipation factor |
0.00001 |
Measured temperature range |
0~125℃ |
Temperature measurement error |
±0.5℃ |
AC Test Voltage (RMS) |
500~2200V Continuously adjustable, frequency 50Hz |
DC Test Voltage |
0~500V Continuously adjustable |
Power consumption |
100W |
Ambient temperature |
0℃~40℃ |
Relative humidity(RH) |
<75% |
Dimensions |
500mm × 360mm × 420mm |
Total weight |
22KG |
4. Functions & Features
1). Oil cup is cleaned automatically and has electrode calibration function.
2). Internal standard capacitor electrode type capacitor, with electrode spacing 2mm.
3). Large-screen LCD, backlight, display is clear. Touch screen operation, more conveniently.
4). More accurate. Testing AC power supply adopts AC - DC - AC conversion method.
5). Equipped with over-voltage, over-current and limit protections.
6). Measured temperature and system clock.
7). Medium frequency induction heating electrode cup and short time uniform heating.
8). Direct measure temperature with probe placed inside the measuring electrode cup.
9). Empty cup self-calibration.
5. Stocks at office
6. Packing, Delivery, Shipping and Payment
7. Certificates
8. Main Accessories
No. |
Item |
Qty |
1 |
Main engine |
1 |
2 |
Oil Cup |
1 |
3 |
Oil tube |
1 |
4 |
Oil cup oil hole |
1 |
5 |
Oil cup observation hole |
1 |
6 |
Dedicated test cable |
1 |
7 |
50ml measuring cylinder |
1 |
8 |
Power line |
1 |
9 |
Fuse pipe |
2 |
10 |
Print paper |
2 |
11 |
Allen wrench(5*30) |
1 |
12 |
Allen wrench(1.5*15) |
1 |
Welcome to visit our website and visit our company, welcome your inquiry and contact.
We are committed to improving the reliability and practicability of Automatic ASTM D924 Transformer Oil Dielectric Loss Tester (TP-6100A), and continue to provide long-term and excellent services to new and old users with a sincere and responsible attitude. All staff of our company hold a meticulous, serious and responsible attitude to meet the needs of customers as the goal, to continue to innovate forward as their own responsibility. There are a wide range of service objects with different expectations and needs. Therefore, good service needs to maintain a kind of flexibility.
Hot Tags: FT-YJS-058A Auto Clean Insulating Oil Dielectric Loss and Resistivity Tester, China, suppliers, manufacturers, customized, price, in stock, buy discount, Transformer direct resistance speed meter, digital high voltage insulation resistance tester, oil density measuring equipment, lube oil filtration, Transformer TTR Tester, oil regeneration plant