We supply dissolved gas analyzer, Transformer Power Factor Tester, resistivity tester to trading companies or domestic market in the past years, and established our reputation with good quality and reasonable price. We constantly strive to maintain the company's code of conduct, improve the level of governance, and lay a solid foundation for becoming an excellent enterprise. After years of development, the company has accumulated a profound cultural heritage and rich cultural connotations. The company has an experienced team, with exquisite craftsmanship, precise design and manufacture, to assist customers in achieving automated production and customized personalized equipment.
1. Introduction & Application
The FUOOTECH's FT-FPC Automatic Closed Flash Point Tester is designed and made as per the National Standard of People's Republic of China GB/T261-2008; GB/T21615-2008; European standard ASTM-D93, it is an advanced analytical instrument, mainly used to determine the closed flash point of petroleum products.
This Automated Closed Cup Flash Point Tester is controlled by microcomputer. It is automatic electronic ignition without the need of gas source. It is reliable and easy to operate with self-test function, high measurement accuracy, good repeatability and stable performance. It is an ideal instrument for closed flash point testing in electrical power, petroleum, chemical, power, environmental protection, railway and scientific research, etc.
2. About Flash Point of Transformer Oil
Flash Point is the lowest temperature at which a liquid can form an ignitable mixture in air near the surface of the liquid. The lower the flash point, the easier it is to ignite the material. Nowadays, Flash point is used to say non-volatile or flammable substances whether exit highly volatile or flammable substances, and has been one of the most important parameters for petroleum products transportation, storage, operations, safety management…
Flash Point of Transformer Oil is defined as the temperature at which light hydrocarbon present in the transformer oil starts evaporating causing a flash on introduction of source under specified condition. It is an important parameter which must be maintained above 140 degree Celsius.
Light hydrocarbons are already present in the transformer oil and gets evaporated under the influence of heat due to electrical stress. Light hydrocarbons are also generated by the degradation of the oil due to aging / heat / electrical stress.
Flash Point of new Transformer Oil should be more than 140 degree Celsius. In any case, this temperature should not be allowed to be less than 130 degree Celsius for transformer oil in service.
3. Technical Parameters
Display |
7-inch color touch LCD display screen |
Test method |
Thermocouple differential detection |
Test range |
Room temperature ~ 400°C |
Repeatability |
≤110°C: 2°C |
≥110°C: 4°C |
|
Accuracy |
When flash point <110℃: ±1℃ |
When flash point ≥110℃: ±2℃ |
|
Reproducibility |
≤8°C |
Resolution |
0.1°C |
Temperature Sensor |
Platinum resistance (PT100) |
Ignition mode |
Automatically electronic ignition |
Printer |
High speed thermal printer |
Flash Fire Sensor |
Ion detection ring |
Cooling mode |
Built-in Forced air cooling |
Power consumption: |
≤350W |
Ambient Temperature |
10°C ~ 45°C |
Relative Humidity |
≤85% |
Power |
AC220V ±10% |
Dimension |
420*220*290mm |
Weight |
7.5 kg |
4. Functions & Features
1). 7-inch industrial color touch screen display, no mechanical contacts, dustproof and waterproof, anti-radio frequency interference, long service life.
2). Equipped with parameter prompt function such as test date and test time.
3). Built-in Germany automatic atmospheric pressure detector: automatic to detect and correct the atmosphere pressure; self-adjust the flash point.
4). Able to store 999 testing data.
5). With imported PT100 sensor for differential detection, automatic correction of system deviation.
6). Cover opening, ignition, detection and printing data are automatically completed.
7). Large-caliber air duct and large air volume fan quickly cools down to shorten the test time.
8). Use built-in battery to ensure that test data can be stored for 5 ~ 8 years even when the external power is off.
9). Built-in high speed thermal printer, quieter, faster and clearer.
5. Stocks at office
6. Packing, Delivery, Shipping and Payment
7. Certificates
8. Accessories
1) Main Tester 1 Set
2) Sample cup 1 Pcs
3) AC power cord 1 Set
4) Printing paper 1 Roll
5) Instruction manual 1 Copy
6) Certificate 1 Set
Welcome to visit our website and visit our company, welcome your inquiry and contact.
All along, our company has been innovating TBT-261A Automatic Pensky-Martens Closed Cup Flash Point Tester, expanding production scale, establishing its own brand, and constantly improving its brand image. We continue to improve the level of technological research and development, accelerate the construction of a global marketing system, and further consolidate our position in the industry We are going to establish a bridge of friendship with you in our good quality products and on the basis of eagerness and credibility, in order to create a great future with you.
Hot Tags: closed cup transformer oil flash point tester, China, suppliers, manufacturers, customized, price, in stock, buy discount, oil purifier unit, transformer oil purification unit, 100A Transformer DC Resistance Tester, oil thermal vacuum purification system, fuel oil filtration, Transformer Oil Acidity Tester