Maintenance and care of automatic trace moisture analyzer
1. Maintenance of electrolyte: Store electrolyte in a well-ventilated place with an ambient temperature below 35°C, no lower limit for low temperature, and a relative humidity of no more than 75%. Replacement of silica gel pad: If the silica gel pad at the sample injection port of the trace moisture analyzer is used for too long, the pinhole passing through the silica gel pad will become non-contractile, allowing moisture in the atmosphere to enter the titration cell and cause measurement errors. At this time, it should be replaced. Replacement of silica gel: When the silica gel in the drying tube changes from blue to light blue or red, new silica gel should be replaced.
2. Maintenance of the ground joint of the titration cell: Rotate the ground joint of the titration cell approximately once a week. If it cannot be rotated easily, carefully remove it, clean the ground joint and the socket, and re-apply a thin layer of vacuum grease. Note: It is not easy to apply too much vacuum grease, otherwise it will enter the titration cell and cause measurement errors. If you do not check it in time, the vacuum grease will harden over time, and the parts connected by the ground joint may stick together and cannot be disassembled. Therefore, they should be maintained regularly to make them easy to disassemble and clean.
3. Treatment of the bonding of the ground joint of the titration cell: If the ground joint of the titration cell is firmly bonded together and difficult to disassemble, follow the procedure: drain the electrolyte in the titration cell and rinse it clean. Inject a small amount of acetone around the ground joint, and then gently turn the parts at the ground joint by hand to disassemble.
If it still cannot be disassembled, please put the titration cell into a 2-liter beaker and slowly add 5% potassium chloride solution to soak it. Be very careful about the liquid level and do not let the lead sleeve ends of the measuring electrode and cathode chamber electrode enter the liquid. After soaking for about ten hours or 24 hours, it can be disassembled. This method can be repeated.
FTWS-809 Automatic Transformer Oil Moisture Tester produced by FUOOTECH® is used for Coulometric water determination according to Karl Fischer as per ASTM D1533. Karl Fischer Coulomb titration is the most reliable method to detect trace moisture for different substances, which was utilized successfully by FTWS-809 trace moisture titrator.
This trace water tester, with its successful application of the method and utilization of computer aided control system, enables a fast determination and higher accuracy while titrated findings are directly displayed on a digital screen. Its self-diagnosis for fault and information display functions help to achieve a more efficient automatic operation.
This Karl Fischer transformer oil water content testing equipment adopts new processor and circuit, excellent low-power consumption. Use measuring electrode signal as the criterion for the end of electrolysis. The stability and accuracy are the critical factor of effect on measurement accuracy. Due to use advanced components and method, make it possible to accurately measure electrode signal and further understand the characteristics of electrolyte. Accurately measure the trace water in liquid, solid and gas, applied to electric power, petroleum, chemical, pharmaceutical and food industries.
If you are interested in any product, please feel free to contact us!